ABL800 FLEX

Parameters

ABL800 FLEX

Type

Para-
meters

Units

Measuring
range

Test
range


pH pH pH scale
nmol/L
6.300-8.000
10.0-501
6.85-7.55
28-141

Blood gas

pCO2

mmHg
kPa
torr

5.0-250
0.67-33.3
5.0-250
17-160
2.27-21.3
17-160

-

pO2

mmHg
kPa
torr
0.0-800
0.00-107
0.0-800
20-580
2.67-77.3
20-580

Electrolytes cCI- mmol/L
meq/L
7-350
7-350
95-150
95-150

- cCa2+ mmol/L
meq/L
mg/dL
0.20-9.99
0.40-19.98
0.80-40.04
0.51-2.2
1.0-4.4
2.0-8.8

- cK+ mmol/L
meq/L
0.5-25.0
0.5-25.0
2-8
2-8

- cNa+ mmol/L
meq/L
7-350
7-350
120-180
120-180

Metabolites cGlu mmol/L
mg/dL
0.0-60
0-1081
0.5-15
9.0-270

- cLac mmol/L
mg/dL
meq/L
0.0-30
0.270
0.0-30
0.5-15
4.5-135
0.5-15

- cCrea µmol/L
mg/dL
10-1800
0.1-20.3
50-1500
0.57-17.0
Find out more about cCrea on the ABL800 FLEX

- ctBil µmol/L
mg/dL
mg/L
0-1000
0.0-58.5
0-585
0-400
0.0-23.4
0-234

Oximetry ctHb

g/dL
mmol/L
g/L

0.00-27.7
0.00-17.2
0.0-277
2.5-23
25-230
1.55-14.2

- sO2 %
Fraction
0.0-100.0
0.000-1.000
0-100
0-1

- FO2Hb %
Fraction
0.0-100.0
0.000-1.000
0-100
0-1

- FCOHb %
Fraction
0.0-100.0
0.000-1.000
0-20
0.0-0.2

- FMetHb %
Fraction
0.0-100.0
0.000-1.000
0-20
0.0-0.2

- FHHb %
Fraction
0.0-100.0
0.000-1.000
0-100
0-1

- FHbF %
Fraction
0.0-100
0.00-1.00
0-80
0.0-0.8


The Measuring range for a parameter is the range within which the analyzer is physically capable of measuring. The measuring range corresponds to the “range of indication” as defined in the “International vocabulary of basic and general terms in the metrology” (VIM).

The Test range for a parameter is the range within which accuracy and precision of a measured parameter has been specified and intended to lie within specified limits. The test range corresponds to the “measuring range” as defined in the “International vocabulary of basic and general terms in the metrology” (VIM).

 
 
 
Services
Our support team is ready to ensure that you get the most out of your solution from day one.
 
In your department
Find out how we can improve workflows in your department.
 
Related info